Engineering Specifications
Multi-Sensor Area-Velocity Flow Meter
Greyline AVMS 5.1

AREA-VELOCITY FLOW MONITOR SPECIFICATIONS

SCOPE: This specification covers an ultrasonic, multi-sensor area-velocity flow monitor as manufactured by Greyline Instruments, Massena, New York / Long Sault, Ontario. The instrument shall provide for indicating, transmitting and totalizing of the flow rate through partially filled or surcharged round pipes and rectangular, trapezoidal, egg or irregular shaped open channels.

A. GENERAL

  1. Flow Monitor to consist of three submersible ultrasonic sensors, connecting cable, and remote enclosures with indicating, transmitting and controlling electronics. The system shall have no moving parts.
  2. Level measurement accuracy shall be ±0.25% of Range. Velocity measurement accuracy shall be ±2% of average reading from multiple sensors.

B. SENSING ELEMENTS

  1. Three ultrasonic sensors shall be rated IP68 for continuous submersion in liquids.
  2. Using the Doppler principle, three sensors shall measure fluid velocities from 0.1 to 20 ft/sec (0.03 to 6.2 m/sec) and reverse flow to -5 ft/sec (-1.5 m/sec). The fluid shall contain bubbles or solids with a minimum size of 100 microns and a minimum concentration of 75 ppm to act as acoustic reflectors.
  3. Readings from three separately mounted velocity sensors shall be continuously averaged by the instrument.
  4. Zero readings shall be automatically factored out if one or more velocity sensors are not submerged or become covered in sediment.
  5. Using ultrasonic echo-ranging principle, one of the submerged sensors shall measure liquid level from 1" to 15 ft. (25.4 mm to 4.57 m). Minimum level shall not be greater than 1" (25.4 mm). Level sensing circuitry shall include a temperature sensor for automatic temperature compensation.
  6. Sensors shall be constructed of PVC and epoxy resin.
  7. Sensor operating temperature shall be 5°F to 150°F (-15°C to 65°C).
  8. Shall include manufacturer's recommended stainless steel sensor mounting brackets.

C. SENSOR CONNECTING CABLE

  1. Provide minimum length 25 ft (7.6 m) tri-coaxial cable with potted bond to each Sensor head. Sensor cable shall be waterproof and electrically shielded. Exposed material shall be polyurethane only.
  2. Extended sensor cable shall be shielded tri-coaxial to a maximum length of 500 ft (152m). Cable shall be spliced with screw terminal connections in manufacturer’s recommended IP66 Junction Box.

D. TRANSMITTER

  1. The transmitter shall provide for field calibration to round pipes and open channels of any shape.
  2. Calibration shall be via built-in 5-key calibration system with menu selection of parameters with English, French or Spanish language selection.
  3. Calibration data shall be password protected and permanently stored through power interruptions for a minimum of 12 months.
  4. Field calibration shall allow selection and automatic conversion of measurement units, measurement span, high/low flow alarm relay and flow proportional relay pulse rates.
  5. Transmitter shall permit field programmable damping to smooth output in turbulent flow conditions.
  6. Transmitter operating temperature shall be from -5° to 140°F (-20° to 60°C). Transmitter shall contain a thermostat-controlled enclosure heater for condensation protection below 30°F (-1°C).
  7. Transmitter shall have three isolated 4-20mA outputs rated for 1000 ohm maximum load with menu-selectable 0-5VDC alternative. Outputs shall be configured to transmit level, velocity and flow.
  8. Provide two relay contacts rated 5 amp SPDT programmable for flow proportionate pulse to a remote totalizer or sampler, high-low flow, velocity and/or level alarm, echo loss alarm.
  9. Provide a white, backlit matrix LCD display indicating flow rate, level, velocity, relay states and 14-digit totalizer in user-selected engineering units.
  10. Transmitter display indicating flow rate, units of calibration, totalizer and relay states shall be visible without opening cover.
  11. Transmitter shall be housed in a wall-mount, watertight NEMA4X (IP66) enclosure with hinged, clear cover. Mounting hardware shall be included.
  12. Transmitter electronics shall be surge protected on AC power input, sensor and 4-20mA outputs.
  13. Transmitter power input shall be 100-240VAC 50-60Hz with power consumption of 5.28 Watts or less.
  14. The transmitter shall permit plug-in field installation and auto-detection of optional accessories including data logger and additional control relays.

E. MULTI-SENSOR TRANSMITTER

  1. The flowmeter shall include a digitally controlled remote transmitter-receiver module with terminal connections for velocity and level sensors.
  2. The sensor transmitter electronics shall be housed in a NEMA4X (IP66) polycarbonate enclosure.
  3. Sensor transmitter electronics shall be powered by connection to the flow meter transmitter.
  4. Enclosure interconnect cable shall be 100 ft (30 m) length heavy-duty Cat 5.

F. OPTIONAL FEATURES FOR INSERTION AS REQUIRED

Sensors
  1. Have a separate non-contacting ultrasonic level sensor mounted above the liquid. The non-contacting level sensor shall be rated for measurement range from 8" to 15 ft. (203.2 mm to 4.5 m) and shall include integral temperature compensation.
  2. Sensor cables shall be 50 ft. (15 m) continuous length with potted bond to the Sensor head.
  3. Sensor cables shall be 100 ft. (30 m) continuous length with potted bond to the Sensor head.
  4. Multi-Sensor Transmitter interconnect cable shall be 328 ft (100 m) length heavy-duty Cat 5.
  5. Separate length extended sensor cable shall be shielded tri-coaxial to a total maximum length of 500 ft (152m). Cable shall be spliced with screw terminal connections in manufacturers recommended IP66 Junction Box.

Electronics

  1. Have a built-in Have a built-in 2 million point Data Logger with USB output to flash drives or mass storage devices. Data logger shall support time and date-stamped logging and generate formatted flow reports including total, average, minimum, maximum and times of occurrence. Include Windows software for data log graphing and export.
  2. Have a thermostat controlled enclosure heater for Transmitter operation at temperatures below freezing.
  3. Have manufacturer's recommended enclosure sunscreen to permit Electronics enclosure mounting in direct sunlight.
  4. Transmitter power input shall be 9-32VDC with minimum power consumption of 6.5 W max.

G. MANUFACTURER

Area-Velocity Flow Monitor shall be Greyline Model AVMS 5.1 and warranted against defects in materials and workmanship for one year.


Specifications are subject to change without notice. 

-- RESOURCES --
QZ02L Sensor Data Sheet
Back to AVMS 5.1 Description
Back to Greyline Instruments Product List


Please contact Greyline if you need more information or for advice in your application. We can provide quotations and refer you to the Greyline sales representative in your area. (To receive complete information please provide your name, mailing address, phone and fax numbers):

Canada:
Greyline Instruments Inc.
16456 Sixsmith Drive
Long Sault, Ont. K0C 1P0
Tel: 613-938-8956 / 888-473-9546
Fax: 613-938-4857
Email:
info@greyline.com
USA:
Greyline Instruments Inc.
105 Water Street
Massena, NY 13662
Tel: 315-788-9500 / 888-473-9546
Fax: 315-764-0419
Email:
info@greyline.com